UEi Test Instruments IRT803

UEi IRT803 Insulation Resistance Tester User Manual

Model: IRT803 | Brand: UEi Test Instruments

1. Einleitung

This manual provides essential information for the safe and effective operation of the UEi IRT803 Digital Insulation Resistance Tester. The IRT803 is a versatile instrument designed for measuring insulation resistance, AC/DC voltage, and continuity. It features a 1000V AC/DC measurement capability, auto-ranging, a backlit display, and a DMM rotary style selector. Key functions include Lock, Compare, and Polarization Index (PI)/Dielectric Absorption Ratio (DAR) tests. Please read this manual thoroughly before use to ensure proper handling and to maximize the instrument's performance.

2. Sicherheitshinweise

Always adhere to safety precautions when using electrical testing equipment. The IRT803 is designed with safety in mind, featuring a CATIV rating up to 600V and CATII up to 1000V, ensuring protection in various jobsite environments. Failure to follow safety instructions can result in electric shock, fire, or damage to the instrument.

  • Verwenden Sie das Testgerät nicht, wenn es beschädigt erscheint oder die Testleitungen beschädigt sind.
  • Stellen Sie sicher, dass die richtige Funktion und der richtige Messbereich ausgewählt sind, bevor Sie Messungen durchführen.
  • Always disconnect power to the circuit under test before connecting or disconnecting test leads for resistance measurements.
  • Tragen Sie geeignete persönliche Schutzausrüstung (PSA), wie z. B. isolierende Handschuhe und Schutzbrille.
  • Do not operate the tester in wet conditions or in the presence of explosive gases or dust.
  • Tauschen Sie die Batterien umgehend aus, sobald die Anzeige für niedrigen Batteriestand erscheint, um genaue Messwerte zu gewährleisten.

3. Lieferumfang

Prüfen Sie beim Auspacken, ob alle unten aufgeführten Artikel vorhanden und unbeschädigt sind:

  • UEi IRT803 Insulation Resistance Tester
  • Silicone test leads with alligator clips (red and black) (ATL57)
  • Insulation test probe with alligator clips (red & black)
  • 4 AA-Batterien
  • Bedienungsanleitung
  • Weiche Tragetasche
Contents of the UEi IRT803 Insulation Resistance Tester kit

Image 3.1: All components included with the UEi IRT803 Insulation Resistance Tester, including the meter, test leads, probes, batteries, manual, and soft carrying case.

4. Produktüberschreitungview

The IRT803 is designed for ease of use and durability. It features an easy-to-read LCD display with a backlight for visibility in various lighting conditions, and a work light to illuminate dark environments. The unit is protected by a rugged rubber boot, which also includes convenient test lead holders for compact storage. An auto power-off function helps conserve battery life.

4.1. Hauptmerkmale und Funktionen

  • 600V AC/DC Measurement: Kann sowohl Wechsel- als auch Gleichstromspannung messentagbis 600V.
  • Isolationswiderstand: Measures resistance from 0.01MΩ up to 2GΩ with test voltages von 50V, 100V, 250V, 500V und 1000V.
  • Kontinuität: Zur Überprüfung der Stromkreisdurchgängigkeit.
  • Compare Reading Function: Allows for comparison against a set reference value.
  • PI/DAR Test: Automatic calculation of Polarization Index and Dielectric Absorption Ratio for advanced insulation diagnostics.
  • A-Hold/Hold: Friert den angezeigten Messwert zur bequemen Aufzeichnung ein.
  • Lock Reading: Locks the test voltage for continuous insulation testing.
  • Rotary Dial Selector: Intuitive selection of measurement functions and test voltages.
  • Protective Rubber Boot: Enhances durability and provides lead storage.
Front view of the UEi IRT803 Insulation Resistance Tester

Bild 4.1: Vorderseite view of the IRT803 showing the LCD display, rotary selector, and function buttons.

Diagram of UEi IRT803 showing functions and features

Image 4.2: Visual representation highlighting the main functions and features of the IRT803.

5. Einrichtung

5.1. Einlegen der Batterie

The IRT803 requires four (4) AA batteries for operation. To install or replace batteries:

  1. Stellen Sie sicher, dass das Testgerät ausgeschaltet ist.
  2. Suchen Sie das Batteriefach auf der Rückseite des Geräts.
  3. Verwenden Sie einen Schraubendreher, um die Abdeckung des Batteriefachs zu öffnen.
  4. Insert the four AA batteries, observing the correct polarity (+/-) as indicated inside the compartment.
  5. Schließen Sie den Batteriefachdeckel sicher.

5.2. Messleitungen anschließen

Connect the test leads to the appropriate input jacks on the tester. For most measurements, the red lead connects to the 'V Ω INSULATION' jack and the black lead connects to the 'COM' (common) jack. Ensure connections are firm and secure.

UEi IRT803 with test leads connected and kickstand extended

Image 5.1: The IRT803 with test leads properly connected to the input jacks.

6. Bedienungsanleitung

6.1. Powering On/Off and Display Backlight

  • To power on the unit, rotate the selector dial from 'OFF' to any desired measurement function.
  • To power off, rotate the selector dial back to 'OFF'. The unit also features auto power-off to save battery life after a period of inactivity.
  • Drücken Sie die WÄHLEN button (often indicated by a light bulb icon) to toggle the display backlight on or off.

6.2. Prüfung des Isolationswiderstands

The IRT803 can perform insulation resistance tests at various voltages.

  1. Select Test Voltage: Rotate the selector dial to the desired insulation test voltage (50 V, 100 V, 250 V, 500 V oder 1000 V).
  2. Messleitungen anschließen: Connect the red test lead to the circuit under test and the black test lead to the ground or common point. Ensure the circuit is de-energized.
  3. Test starten: Press and hold the large red PRÜFEN button to begin the insulation test. The test voltage will be applied, and the resistance reading will be displayed.
  4. Sperrfunktion: Für kontinuierliche Tests drücken Sie die Taste SPERREN button after initiating the test. This will maintain the test voltage without needing to hold the PRÜFEN Taste. Drücken Sie PRÜFEN erneut zum Entsperren.
  5. Vergleichsfunktion: Verwenden Sie die VERGLEICH button to set a reference value and compare subsequent readings against it.
  6. PI/DAR Test: Select the PI/DAR function on the dial. The meter will automatically perform the Polarization Index and Dielectric Absorption Ratio tests, displaying the calculated values.
  7. Read and Record: Observe the reading on the LCD. Use the HALTEN button to freeze the display for easier recording.
  8. Entladung: After completing an insulation test, the circuit under test should be discharged. The IRT803 typically handles this automatically, but always exercise caution.
UEi IRT803 in use, measuring insulation resistance on electrical wiring

Image 6.1: The IRT803 being used to measure insulation resistance on electrical wiring.

Close-up of test leads connected to the UEi IRT803

Bild 6.2: Eine Nahaufnahme view of the test leads securely connected to the IRT803's input terminals.

6.3. Volumentage Prüfung (AC/DC)

Zur Messung von AC- oder DC-Voltage up to 600V:

  1. Rotate the selector dial to the 'V' position.
  2. Connect the red test lead to the positive or live point and the black test lead to the negative or neutral/ground point of the circuit.
  3. Das Display zeigt das gemessene Volumen an.tage. The meter will automatically detect AC or DC voltage.

6.4. Durchgangsprüfung

To check for continuity in a circuit:

  1. Rotate the selector dial to the 'Ω' (Ohms) position, which typically includes continuity.
  2. Connect the test leads across the component or circuit path you wish to test.
  3. If there is continuity (low resistance), the meter will typically emit an audible beep and display a low resistance value.

7. Wartung

7.1. Reinigung

To maintain the accuracy and longevity of your IRT803:

  • Wischen Sie das casing mit Adamp Tuch und mildem Reinigungsmittel. Keine Scheuer- oder Lösungsmittel verwenden.
  • Ensure the unit is completely dry before storage or next use.

7.2. Batteriewechsel

Replace batteries as described in Section 5.1 when the low battery indicator appears on the display. Always use fresh AA alkaline batteries.

7.3. Speicherung

When not in use, store the IRT803 in its soft carrying case in a cool, dry environment, away from direct sunlight and extreme temperatures. If storing for extended periods, remove the batteries to prevent leakage.

8. Fehlerbehebung

If you encounter issues with your IRT803, refer to the following common problems and solutions:

ProblemMögliche UrsacheLösung
Tester does not power onLeere oder falsch eingelegte BatterienBatteriepolung prüfen; durch neue AA-Batterien ersetzen.
Keine AnzeigeIncorrect function selected; open circuit; poor test lead connectionVerify selector dial position; ensure test leads are firmly connected to the circuit and the meter; check for breaks in the circuit.
Ungenaue MesswerteLow battery; environmental interference; incorrect range/functionReplace batteries; move away from strong electromagnetic fields; ensure correct function and test voltage werden ausgewählt.
Test leads not workingDamaged leads; poor connectionInspect leads for damage; ensure they are fully inserted into the jacks. Replace if damaged.

If the problem persists after attempting these solutions, please contact UEi Test Instruments customer support.

9. Spezifikationen

ParameterWert
ModellnummerIRT803
MarkeUei-Testgeräte
Isolationstest Bdtages50 V, 100 V, 250 V, 500 V, 1000 V
Isolationswiderstandsbereich0.01MΩ to 2GΩ
Insulation Resistance Accuracy±1.5-3% + 5 digits (0.01MΩ-500MΩ)
Bandtage MessungBis zu 600 V AC/DC
Kontinuitätsbereich20.00 Ω to 20.0 kΩ (as per product description)
SicherheitsbewertungCATIV 600V, CATII 1000V
Stromquelle4 x AA-Batterien
AnzeigeLCD mit Hintergrundbeleuchtung
Technische Daten2.7 x 3.52 x 8.27 Zoll
Artikelgewicht1.8 Pfund
FarbeGelb

10. Garantieinformationen

The UEi IRT803 Insulation Resistance Tester is backed by a 1 Jahr eingeschränkte Garantie from the date of purchase. This warranty covers defects in materials and workmanship under normal use. It does not cover damage resulting from misuse, unauthorized modification, accident, or neglect. For specific terms and conditions, please refer to the warranty card included with your product or contact UEi Test Instruments customer support.

Certificate of Conformity for UEi Test Instruments product

Abbildung 10.1: Beispielample of a Certificate of Conformity, indicating adherence to quality standards.

11. Unterstützung

For technical assistance, troubleshooting beyond this manual, or warranty claims, please contact UEi Test Instruments customer support. You can find more information and contact details by visiting the official UEi Test Instruments webWebsite oder deren Amazon-Shopseite.

UEi Test Instruments Store: Besuchen Sie den Shop

© 2025 UEi Test Instruments. All rights reserved.

Ask a question about this manual