1. Einleitung
This manual provides detailed instructions for the safe and effective operation of the WOOXGEHM T6125 Digital Insulation Resistance Tester. The T6125 is a robust, shockproof, and dustproof instrument designed for measuring insulation resistance, absorption ratio (DAR), polarization index (PI), AC voltage, and low resistance. It features a wide measurement range, multiple output voltage levels, a large LCD with backlight, and strong anti-interference capabilities, making it suitable for various electrical equipment testing applications.

Figure 1: WOOXGEHM T6125 Digital Insulation Resistance Tester.
2. Sicherheitshinweise
To ensure safe operation and prevent damage to the instrument or injury to personnel, please read and understand all safety instructions before use.
- Beachten Sie stets die lokalen und nationalen Sicherheitsvorschriften.
- Das Testgerät darf nicht bedient werden, wenn es beschädigt erscheint oder nicht ordnungsgemäß funktioniert.
- Ensure the circuit under test is de-energized and discharged before connecting test leads for resistance measurements.
- Verwenden Sie ausschließlich die mitgelieferten Messleitungen und Zubehörteile.
- Avoid touching the test leads or the circuit under test during measurement, especially when high voltages sind vorhanden.
- Do not use the instrument in wet environments or explosive atmospheres.
- Tauschen Sie die Batterien umgehend aus, sobald die Anzeige für niedrigen Batteriestand erscheint.
3. Packungsinhalt
Bitte überprüfen Sie, ob alle unten aufgeführten Artikel in Ihrem Paket enthalten sind:
- 1 x WOOXGEHM T6125 Digital Insulation Resistance Tester
- 1 x Paar Messleitungen
- 1 x External Power Plug (5.5mm)
- 2 x Krokodilklemmen
- 1 x Stoffbeutel
- 1 x Benutzerhandbuch (Englisch)
4. Produktüberschreitungview
The T6125 insulation tester is designed for comprehensive electrical insulation testing. Key features include:
- Isolationswiderstandsmessung: Measures insulation opposition, absorption ratio, and polarization index.
- AC-Lautstärketage-Messung: Kann Wechselspannung messentage.
- Low Resistance Measurement: For continuity checks.
- Mehrfachtest Voltages: Bietet Testspannungen von 100 V, 250 V, 500 V, 1000 V und 2500 V an.tage-Optionen.
- Automatische Entladefunktion: Ensures safety after high voltage testet.
- Datenaufnahme: Friert den angezeigten Messwert zur einfacheren Aufzeichnung ein.
- Over-range Alarm: Alerts the user when measurements exceed the selected range.
- Automatisches Herunterfahren: Schaltet sich nach 30 Minuten Inaktivität automatisch ab, um die Batterielebensdauer zu verlängern.
- Großes, hintergrundbeleuchtetes LCD-Display: Liefert klare Messwerte unter verschiedenen Lichtverhältnissen.
- Langlebiges Design: Shockproof, dustproof, and dampness-proof construction.

Figure 2: Front panel layout and key components of the T6125.
- Hohe Lautstärketage Output Black Negative Jack: Connects the negative test lead for high voltage testet.
- Hohe Lautstärketage Output Red Positive Jack: Connects the positive test lead for high voltage testet.
- Resistor Red Positive Jack: Connects the positive test lead for resistance measurements.
- Resistive Black Negative Jack: Connects the negative test lead for resistance measurements.
- 12V-Stromeingangsbuchse: Für externe Stromversorgung.
- LCD Anzeige: Shows measurement readings and status.
- Test-Taste: Startet eine Messung.
- Datenhaltetaste: Freezes the current display.
- Taste für Hintergrundbeleuchtung: Aktiviert/deaktiviert die LCD-Hintergrundbeleuchtung.
- DAR (Absorption Ratio) / PI (Polarization Index) Button: Selects DAR or PI measurement mode.
- Gangdrehschalter: Selects measurement function and test voltage.

Figure 3: The T6125 features an automatic shutdown function after 30 minutes of inactivity to conserve battery life.
5. Einrichtung
5.1 Einlegen der Batterie
The T6125 requires 8 x 1.5V AA batteries (not included).
- Suchen Sie das Batteriefach auf der Rückseite des Geräts.
- Öffnen Sie die Abdeckung des Batteriefachs.
- Legen Sie 8 AA-Batterien ein und achten Sie dabei auf die richtige Polarität (+/-).
- Schließen Sie den Batteriefachdeckel fest.
5.2 Anschließen der Testleitungen
Connect the test leads to the appropriate jacks on the front panel based on the measurement type. For insulation resistance tests, use the High Voltage Output jacks. For low resistance, use the Resistor jacks.
6. Bedienungsanleitung
6.1 Ein-/Ausschalten
Rotate the Gear Rotary Switch from the "OFF" position to any measurement function to power on the device. Rotate back to "OFF" to power off. The device will automatically shut down after 30 minutes of inactivity.
6.2 Isolationswiderstandsmessung
This function measures the insulation resistance of electrical equipment.
- Stellen Sie sicher, dass der zu prüfende Stromkreis spannungsfrei und sicher entladen ist.
- Connect the test leads to the High Voltage Output jacks (red to positive, black to negative).
- Connect the other ends of the test leads to the insulation points of the equipment to be tested.
- Rotate the Gear Rotary Switch to the desired insulation test voltage (e.g., 100V, 250V, 500V, 1000V, 2500V).
- Halten Sie die PRÜFEN button. The measurement will begin, and the resistance value will be displayed on the LCD.
- Lassen Sie den PRÜFEN button to stop the measurement. The instrument will automatically discharge the circuit.
- Um den Messwert zu speichern, drücken Sie die Taste HALTEN button during or after a measurement. Press again to release.

Figure 4: The T6125 offers a wide insulation test measurement range up to 2500V and 200.0GΩ.
6.3 Messung des Absorptionsverhältnisses (DAR) und des Polarisationsindex (PI)
DAR and PI are used to assess the quality of insulation over time.
- Follow steps 1-3 for Insulation Resistance Measurement.
- Rotate the Gear Rotary Switch to the desired insulation test voltage.
- Drücken Sie die DAR/PI button to select the desired mode. The display will indicate "DAR" or "PI".
- Halten Sie die PRÜFEN button. The instrument will automatically perform the timed measurements and display the DAR or PI value.
- Lassen Sie den PRÜFEN Taste.
6.4 AC-Voltage Messung
Zur Messung von AC-Voltage:
- Rotate the Gear Rotary Switch to the "ACV" position.
- Connect the test leads to the circuit where AC voltage muss gemessen werden.
- Die AC-Voltage wird auf dem LCD angezeigt.
6.5 Messung niedriger Widerstände
For continuity or low resistance checks:
- Rotate the Gear Rotary Switch to the "Ω" (low resistance) position.
- Connect the test leads to the Resistor jacks.
- Connect the other ends of the test leads to the component or circuit to be measured.
- Der Widerstandswert wird auf dem LCD-Display angezeigt.
7. Wartung
7.1 Reinigung
Wischen Sie das Instrument mit Anzeige abamp cloth and mild detergent. Do not use abrasive cleaners or solvents. Ensure the instrument is completely dry before storage or next use.
7.2 Speicherung
When not in use for extended periods, remove the batteries to prevent leakage. Store the instrument in the provided cloth bag in a cool, dry place, away from direct sunlight and extreme temperatures.
- Lagertemperatur: -20 °C bis 60 °C
- Working Conditions: 0℃ to 40℃
- Relative Luftfeuchtigkeit: 40 % bis 75 %
8. Fehlerbehebung
If you encounter issues with your WOOXGEHM T6125, refer to the following common problems and solutions:
- Kein Strom:
- Prüfen Sie, ob die Batterien richtig eingesetzt sind und ausreichend geladen sind. Tauschen Sie sie gegebenenfalls aus.
- Ensure the Gear Rotary Switch is not in the "OFF" position.
- Ungenaue Messwerte:
- Verify test leads are securely connected and not damaged.
- Ensure the correct measurement function and test voltage werden ausgewählt.
- Clean the test points on the equipment being measured.
- Prüfen Sie auf externe Störquellen.
- "OL" or Over-range Display:
- The measured value exceeds the selected range. Try a higher test voltage or a different range if applicable.
- Ensure proper connection to the circuit.
Wenn das Problem weiterhin besteht, wenden Sie sich an den Kundendienst.
9. Spezifikationen
| Modell | T6125 |
| Material | ABS |
| Stromversorgung | 8 x 1.5 V AA-Batterien (nicht im Lieferumfang enthalten) |
| Artikelgröße | 179 x 133 x 70 mm (7.04 x 5.23 x 2.75 Zoll) |
| Artikelgewicht | 570 g |
| Lagertemperatur | -20℃ bis 60℃ |
| Arbeitsbedingungen | 0℃ bis 40℃ |
| Höhe | 2000 m |
| Relative Luftfeuchtigkeit | 40 % bis 75 % |
Insulation Test Accuracy (T-6125)
| Ausgangsvolumentage | Reichweite | Auflösung | Maximaler Prüfstrom | Genauigkeit |
|---|---|---|---|---|
| 250 V (± 10%) | 0 bis 20 MΩ | 0.01 MΩ | 2 mA | ± (3 % + 5) |
| 20 bis 200 MΩ | 0.1 MΩ | ± (3 % + 5) | ||
| 0.2G~2GΩ | 0.001 GΩ | ± (10 % + 5) | ||
| 500 V (± 10%) | 0 bis 20 MΩ | 0.01 MΩ | 2 mA | ± (3 % + 5) |
| 20 bis 200 MΩ | 0.1 MΩ | ± (3 % + 5) | ||
| 0.2G~2GΩ | 0.001 GΩ | ± (10 % + 5) | ||
| 1000 V (± 10%) | 0 bis 20 MΩ | 0.01 MΩ | 2 mA | ± (3 % + 5) |
| 20 bis 200 MΩ | 0.1 MΩ | ± (3 % + 5) | ||
| 0.2G~2GΩ | 0.001 GΩ | ± (10 % + 5) | ||
| 2500 V (± 10%) | 0 bis 20 MΩ | 0.01 MΩ | 2 mA | ± (3 % + 5) |
| 20 bis 200 MΩ | 0.1 MΩ | ± (5 % + 5) | ||
| 0.2G~2GΩ | 0.001 GΩ | ± (10 % + 5) | ||
| 2G~20GΩ | 0.01 GΩ | ± (10 % + 5) | ||
| AC-Lautstärketage Genauigkeit | Ranging: 600V | Auflösung: 1V | N / A | ± (1.5 % + 5) |
| Widerstandsgenauigkeit | Ranging: 60Ω | Auflösung: 0.01 Ω | N / A | ± (1.2 % + 5) |

Figure 5: Detailed insulation test accuracy specifications for the T6125 model.

Figure 6: Physical dimensions of the WOOXGEHM T6125.
10. Garantie und Support
For warranty information or technical support, please refer to the contact details provided with your purchase or visit the official WOOXGEHM webBewahren Sie Ihren Kaufbeleg als Kaufnachweis auf.