ASCOM NPTX2-BFKAA2

Ascom Teleassistance NPTX2-BFKAA2 Benutzerhandbuch

Model: NPTX2-BFKAA2 | Brand: ASCOM

1. Produktüberschreitungview

The Ascom TeleCARE NPTX2-BFKAA2 is a white radio call receiving unit designed for patient and nurse communication in healthcare environments. It features three distinct buttons with integrated indicator lights and a pull cord for initiating various types of calls or indicating presence.

Ascom Teleassistance NPTX2-BFKAA2 unit with red, yellow, and green buttons, and a red pull cord.

Abbildung 1: Vorderseite view of the Ascom Teleassistance NPTX2-BFKAA2 unit, showing the three colored buttons and the integrated pull cord.

2. Komponenten

The NPTX2-BFKAA2 unit is equipped with the following key components:

  • Red Button with Nurse Symbol: This button, accompanied by an indicator light and a silkscreened nurse symbol, is primarily used for initiating an emergency or direct nurse call. It also features an integrated pull cord for activation.
  • Grüner Knopf: This button, with its own indicator light, is used for functions such as call cancellation or indicating the presence of a nurse.
  • Gelber Knopf: Equipped with an indicator light, this button is designated for requesting general assistance.
  • Screw Terminal Block (Input): Provided for connecting a wired ASCOM teleCARE call unit, allowing integration with existing wired systems.
  • Terminal Block Output: Used for connecting a 2-light signaling window, which provides visual alerts.

3. Einrichtung

Proper installation and programming are crucial for the correct functioning of the NPTX2-BFKAA2 unit. It is recommended that setup be performed by qualified personnel.

  1. Stromversorgung: The unit requires either a local 24 VDC power supply or a 9 V battery. Ensure the power source meets the specified voltage-Anforderungen.
  2. Kabelverbindungen: If integrating with a wired ASCOM teleCARE system, connect the wired call unit to the designated screw terminal block. For visual signaling, connect a 2-light signaling window to the terminal block output.
  3. Programmierung: PROG PORT programming is mandatory for the unit to function correctly within your system. If you are not equipped for this, contact your ASCOM service provider or a certified technician.
  4. Montage: Securely mount the unit in an accessible location, ensuring the pull cord is within reach for its intended users.

4. Bedienungsanleitung

The NPTX2-BFKAA2 unit provides intuitive controls for various call and status functions:

  • Initiating a Nurse Call (Red Button/Pull Cord):
    • To call a nurse, press the red button firmly or pull the red cord downwards.
    • The indicator light on the red button will illuminate, confirming the call has been initiated.
    • An audible and/or visual alert will be sent to the central monitoring system or nurse station.
  • Cancelling a Call / Indicating Nurse Presence (Green Button):
    • To cancel an active call or to indicate that a nurse is present in the room, press the green button.
    • The indicator light on the green button will illuminate, and the red button's light may extinguish, depending on system configuration.
  • Requesting Assistance (Yellow Button):
    • For general assistance that is not an emergency, press the yellow button.
    • The indicator light on the yellow button will illuminate, signaling an assistance request.
  • LED-Anzeigen: Observe the indicator lights on each button. A lit LED typically signifies an active call or status. Refer to your system's specific programming for detailed light patterns.

5. Wartung

Regular maintenance ensures the longevity and reliable operation of your NPTX2-BFKAA2 unit.

  • Reinigung: Clean the unit periodically with a soft, damp cloth. Do not use abrasive cleaners, solvents, or excessive moisture. Ensure the unit is powered off before cleaning.
  • Batteriewechsel: If the unit is powered by a 9V battery, replace it when the low battery indicator (if available and configured) illuminates, or as part of a routine maintenance schedule. Refer to the installation guide for battery access.
  • Inspektion: Regularly inspect the pull cord for any signs of wear or damage. Ensure buttons operate smoothly and lights illuminate correctly.

6. Fehlerbehebung

If you encounter issues with your NPTX2-BFKAA2 unit, consider the following common problems and solutions:

ProblemMögliche UrsacheLösung
Unit does not respond / No lightsKein Strom oder schwacher Akku.Check 24VDC power supply connection or replace 9V battery.
Call initiated but not received at nurse stationRadio interference, unit out of range, or programming issue.Verify unit is within range. Contact a technician for programming verification or radio signal strength check.
Buttons feel stuck or unresponsiveDirt/debris accumulation or mechanical fault.Gently clean around buttons. If issue persists, contact service.
Pull cord is damagedPhysischer Verschleiß.Contact ASCOM service for replacement of the pull cord assembly.

7. Spezifikationen

BesonderheitDetail
ModellnummerNPTX2-BFKAA2
MarkeASCOM
Stromversorgung24 VDC or 9 V Battery
GewichtCa. 227 g (0.5 lbs)
FarbeWhite (Unit shown in image) / Noir (Listed in some specifications, may refer to internal components or a different variant)
Anzahl der Tasten3 (Red, Green, Yellow)
Kontrollleuchten3 (One for each button)
KonnektivitätRadio, Wired ASCOM teleCARE input, 2-light signaling window output
ProgrammierungPROG PORT (Mandatory)

8. Garantie und Support

For information regarding product warranty, technical support, or service, please contact your authorized ASCOM distributor or the point of purchase. Keep your purchase receipt and product serial number handy when contacting support.

ASCOM provides comprehensive support for its healthcare communication solutions. For specific regional contact details, please visit the official ASCOM webWebsite.

Zugehörige Dokumente - NPTX2-BFKAA2

Vorview Ascom teleCARE IP Installationsanleitung
Ausführliche Installationsanleitung für das Ascom teleCARE IP-System mit detaillierten Informationen zu Komponenten wie Raumdisplays (NIRD), Korridordisplays (NICD) und drahtlosen Funktionen für Einrichtungen des betreuten Wohnens und der Pflege.
Vorview Installationsanleitung für das drahtlose Zugschnurmodul Ascom NUWPC3
Detaillierte Installationsanleitung für das Ascom NUWPC3 Wireless Pull Cord Module mit Kabeleinstellung, Batterieinstallation, Modulentfernung und Tasteneinsatzwechsel. Enthält Schritt-für-Schritt-Anleitungen und Sicherheitshinweise.
Vorview Installationsanleitung für das drahtlose Zugschnurmodul Ascom NUWPC3 | Sicherheits- und Montagehinweise
Installationsanleitung für das Ascom NUWPC3 Wireless Pull Cord-Modul. Dieses Dokument enthält Anweisungen zur Montage, Sicherheitsvorkehrungen und Informationen zur Einhaltung gesetzlicher Vorschriften für das Gerät, das Teil des teleCARE-Systems ist.
Vorview Ascom Gateway Benutzerhandbuch v9.0 – Patientenüberwachung und Datenerfassung
Ausführliches Benutzerhandbuch für die Ascom Gateway-Anwendung mit detaillierten Informationen zu Installation, Konfiguration, Integration von Wearables, Erfassung von Vitalparametern und Benutzer-Workflows für medizinisches Fachpersonal.
Vorview Ascom Gateway Benutzerhandbuch: Installation, Konfiguration und Arbeitsabläufe
Ausführliches Benutzerhandbuch für das Ascom Gateway mit detaillierten Informationen zu Installation, Konfiguration, Kopplung von Wearables, Datenerfassung und Benutzer-Workflows für Anwendungen im Gesundheitswesen.
Vorview Sicherheitshinweise und behördliche Informationen zum Ascom Myco 3-Mobilteil
Dieses Dokument enthält wichtige Sicherheitshinweise, Informationen zum Verwendungszweck und zur Einhaltung gesetzlicher Vorschriften für das Ascom Myco 3-Mobilteil. Es umfasst Vorsichtsmaßnahmen, Umweltanforderungen und Erklärungen für verschiedene Regionen, darunter EU/EFTA, Großbritannien, USA und Kanada.