Elitech RC-5+ TE

Elitech RC-5+ TE Digital USB Temperature Data Logger Instruction Manual

Brand: Elitech | Model: RC-5+ TE

1. Einleitung

The Elitech RC-5+ TE is a versatile digital USB temperature data logger designed for reliable and accurate temperature monitoring across various applications, including life sciences, food cold chain, and industrial fields. This device features both an internal and an external sensor, allowing for flexible measurement options. It generates automatic PDF reports without requiring special software, simplifying data access and analysis. The RC-5+ TE is compliant with FDA 21 CFR Part 11 standards, ensuring high quality and regulatory adherence.

Elitech RC-5+ TE Reusable Temperature Data Logger

Image: Elitech RC-5+ TE Reusable Temperature Data Logger, ideal for cold chain transportation temperature monitoring.

Hauptmerkmale:

  • Doppelsensoren: Equipped with both internal and external temperature probes for comprehensive monitoring.
  • Automatic PDF Report: Generates PDF reports directly upon USB connection, no proprietary software needed for basic data access.
  • Hohe Genauigkeit: Provides precise temperature readings with a low margin of error.
  • Large Data Capacity: Stores up to 32,000 data points.
  • Großer Temperaturbereich: Suitable for extreme cold and heat, from -40°C to +85°C (-40°F to 185°F) with the external probe.
  • IP67 Wasserdicht: Offers robust protection against water and dust.
  • Konfigurierbare Parameter: Allows setting of temperature units, alarm ranges, and logging intervals.
  • Lange Akkulaufzeit: Up to 1 year of operation on a single CR2032 battery.

Produkt überview Video:

Video: Ein überview of the Elitech RC-5+ and RC-5+ TE temperature data loggers, highlighting their features and applications.

2. Lieferumfang

Upon opening your Elitech RC-5+ TE package, you should find the following items:

  • Elitech RC-5+ TE Data Logger
  • USB External Probe/Sensor
  • Kalibrierzertifikat
  • Bedienungsanleitung
Elitech RC-5+ TE Package Contents

Image: The Elitech RC-5+ TE package includes the data logger, external probe, calibration certificate, and user manual.

3. Einrichtung

3.1 Einlegen der Batterie

The RC-5+ TE uses a CR2032 lithium metal battery. To install or replace the battery:

  1. Suchen Sie das Batteriefach auf der Rückseite des Datenloggers.
  2. Use a coin or small tool to twist and open the battery cover.
  3. Legen Sie die CR2032-Batterie so ein, dass die positive Seite (+) nach oben zeigt.
  4. Bringen Sie die Batterieabdeckung wieder an und drehen Sie sie fest.
Elitech RC-5+ TE Battery Installation

Image: Step-by-step guide for installing the CR2032 battery into the Elitech RC-5+ TE data logger.

3.2 Softwareinstallation und Erstkonfiguration

While the RC-5+ TE can generate basic PDF reports without software, the Elitechlog software (available for Windows and Mac) provides advanced configuration options and detailed data analysis. To download and install the software:

  1. Besuchen Sie die offizielle Elitech-Website. webWebsite: www.elitechustore.com.
  2. Navigate to the 'Download' section and select 'Software'.
  3. Choose the appropriate software version for your operating system (Windows or Mac) and download the installer.
  4. Folgen Sie den Anweisungen auf dem Bildschirm, um die Installation abzuschließen.
  5. Once installed, connect the RC-5+ TE to your computer's USB port.
  6. Open the Elitechlog software. Go to the 'Settings' tab (or 'Parameter' tab on Windows) to configure the device.
  7. Set the desired logging interval (from 10 seconds up to 12 hours), temperature unit (°C/°F), time zone, and alarm settings.
  8. For the RC-5+ TE, select whether to use the internal or external probe under 'Probe Type'.
  9. Click 'Save' to apply your settings to the device.
  10. Safely eject the device from your computer.

Software Setup and Configuration Video:

Video: Detailed instructions on how to download, install, and configure the Elitechlog software for RC-5+ and RC-5+ TE data loggers.

3.3 Connecting the External Sensor

The RC-5+ TE model includes an external sensor for more precise or remote temperature measurements. Simply plug the external sensor cable into the designated port on the data logger. Ensure it is securely connected.

Elitech RC-5+ TE External Sensor and USB Port

Image: The Elitech RC-5+ TE with its external sensor connected and the USB port exposed.

4. Bedienung

4.1 Starten einer Aufnahme

To begin temperature logging, press and hold the left play button (►) for approximately 5 seconds. The play sign will appear on the LCD, indicating that recording has started.

4.2 Beenden einer Aufnahme

To stop temperature logging, press and hold the right pause button (⏸) for approximately 5 seconds. The stop sign will appear on the LCD, indicating that recording has ceased.

4.3 Viewing Data on the LCD

Drücken Sie die left play button (►) briefly to cycle through the following display menus:

  • Aktuelle Temperatur
  • Anzahl der protokollierten Messwerte
  • Aktuelle Zeit
  • Aktuelles Datum
  • Höchsttemperatur gemessen
  • Minimale gemessene Temperatur
Elitech RC-5+ TE LCD Interface Display

Image: Visual representation of the Elitech RC-5+ TE LCD interface, showing various data points like temperature, logging points, time, date, and min/max values.

4.4 Markierungsereignisse

During logging, you can mark specific events by double-clicking the right pause button (⏸). This feature is useful for noting significant occurrences during the monitoring period.

Elitech RC-5+ TE Event Marking Feature

Image: Instructions on how to clear data and restart the recorder, and how to double-click to mark an event on the Elitech RC-5+ TE.

5. Datenverwaltung

5.1 Automatische PDF-Berichtserstellung

The RC-5+ TE automatically generates a PDF report when plugged into a computer's USB port. No special software is required to view this report.

  1. Ensure the data logger has stopped recording.
  2. Remove the protective cap from the USB connector.
  3. Plug the data logger into an available USB port on your computer.
  4. The device will appear as a removable drive. Open the drive to find the generated PDF report.
Elitech RC-5+ TE Automatic PDF Report Generation

Image: The Elitech RC-5+ TE connected to a laptop, demonstrating automatic PDF and CSV report generation without additional software.

5.2 Exporting Data with Elitechlog Software

For more detailed analysis and export options (e.g., Excel, AELT), use the Elitechlog software:

  1. Connect the data logger to your computer and open the Elitechlog software.
  2. The software will automatically read the data from the device.
  3. Click on the 'Export Data' option within the software interface.
  4. Wählen Sie Ihren Wunsch file format (Excel, PDF, AELT) and specify a save location.
  5. Click 'Save' to export the data.
Elitech RC-5+ TE Software Data Comparison

Image: The Elitechlog software interface showing data comparison graphs and tables from multiple loggers, allowing for powerful analysis.

6. Wartung

6.1 Batteriewechsel

Refer to section 3.1 for detailed instructions on how to replace the CR2032 battery. It is recommended to replace the battery annually or when the low battery indicator appears on the LCD.

6.2 Reinigung

To clean the data logger, simply wipe it with a clean, damp cloth. Avoid using abrasive cleaners or solvents that could damage the device. The device is freezer safe.

6.3 Wasserdichter Schutz

The RC-5+ TE features an IP67 waterproof rating, meaning it is protected against dust ingress and immersion in water up to 1 meter for 30 minutes. Ensure the USB cap is securely in place to maintain its waterproof integrity.

Elitech RC-5+ TE IP67 Waterproof Rating

Image: The Elitech RC-5+ TE data logger submerged in water, demonstrating its IP67 waterproof capability.

7. Fehlerbehebung

7.1 No Data Recorded / "Battery Too Low" Error

  • Check the battery level on the LCD display. If low, replace the CR2032 battery (refer to section 3.1).
  • Ensure the recording was properly started by pressing and holding the left play button until the play sign appears.
  • Verify that the logging interval and start mode are correctly configured in the Elitechlog software.

7.2 "nC" Error (No Connection)

  • This error typically indicates a problem with the external sensor connection. Ensure the external sensor is firmly plugged into the data logger's port.
  • If the issue persists, try unplugging and re-plugging the sensor.
  • If using the external sensor, ensure the 'Probe Type' setting in the Elitechlog software is set to 'External'.

7.3 Software Not Recognizing Device

  • Ensure the Elitechlog software is correctly installed and updated to the latest version.
  • Try connecting the data logger to a different USB port on your computer.
  • Restart your computer and the Elitechlog software.

8. Spezifikationen

BesonderheitSpezifikation
ModellnameRC-5+ TE
Temperature Measurement Range (External Probe)-40°C bis +85°C (-40°F bis 185°F)
Temperature Measurement Range (Internal Probe)-30°C bis +70°C (-22°F bis 158°F)
Temperaturgenauigkeit±0.5°C/±0.9°F (-20°C to +40°C); ±1°C/±1.8°F (others)
Auflösung0.1 °C/°F
Datenkapazität32,000 Punkte
ProtokollierungsintervallKonfigurierbar von 10 Sekunden bis 12 Stunden
AlarmeinstellungenUp to 5 alarm settings
KonnektivitätUSB
WasserdichtigkeitIP67
Akku-TypCR2032 Lithium-Metallbatterie (im Lieferumfang enthalten)
AkkulaufzeitBis zu 1 Jahr
Maße3.2 x 1.3 x 0.5 Zoll
Artikelgewicht2.39 Unzen
MaterialPlastik
EinhaltungFDA 21 CFR Teil 11

9. Garantie und Support

9.1 Garantieinformationen

The Elitech RC-5+ TE Digital USB Temperature Data Logger comes with a 12 Monate Garantie Diese Garantie gilt ab Kaufdatum und deckt Herstellungsfehler ab. Sie gewährleistet, dass das Produkt die spezifizierten Leistungsstandards erfüllt. Bitte bewahren Sie Ihren Kaufbeleg für Garantieansprüche auf.

9.2 Kundendienst

For any technical assistance, troubleshooting, or inquiries regarding your Elitech product, please contact Elitech Customer Support. They offer 24/7 support via email and phone.

  • E-Mail-Support: Refer to the Elitech website or your product packaging for the specific email address.
  • Telefonischer Support: Refer to the Elitech website or your product packaging for the specific phone number.
  • Online-Ressourcen: Besuchen Sie Elitech webSeite? ˅ (www.elitechustore.com) for FAQs, software downloads, and additional user manuals.

Zugehörige Dokumente - RC-5+ TE

Vorview Elitech RC-5+ Temperaturdatenlogger: Benutzerhandbuch und Spezifikationen
Dieses Handbuch enthält detaillierte Anweisungen zur Verwendung des Elitech RC-5+ USB-Temperaturdatenloggers. Es umfasst Einrichtung, Bedienung, Datendownload und Analyse mit der Elitech-Software. Funktionen und technische Daten sind ebenfalls enthalten.
Vorview Elitech RC-51 Mehrzweck-Temperaturdatenlogger – Technische Spezifikationen
Entdecken Sie den Elitech RC-51, einen vielseitigen Temperaturdatenlogger, der für Genauigkeit und Zuverlässigkeit entwickelt wurde. Dieses Dokument enthält technische Spezifikationen, Funktionen und Informationen zum Software-Download für das Modell RC-51.
Vorview Elitech RC-5 Kurzanleitung: Temperaturdatenlogger
Kurzanleitung für den Elitech RC-5 Temperaturdatenlogger mit Informationen zum Einlegen der Batterie, zur Einrichtung der Software, zum Betrieb des Loggers, zu technischen Daten und zur Fehlerbehebung.
Vorview Elitech RC-51 Datenlogger Benutzerhandbuch
Benutzerhandbuch für den Temperaturdatenlogger Elitech RC-51 mit detaillierten Informationen zu Funktionen, Spezifikationen, Einrichtung, Bedienung, Datenaufzeichnung und Batteriewechsel.
Vorview Elitech RC-51 Mehrzweck-Temperaturdatenlogger – Benutzerhandbuch
Ausführliches Benutzerhandbuch für den Elitech RC-51 Mehrzweck-Temperaturdatenlogger, das das Produkt umfassend abdecktview, technische Spezifikationen, Bedienungsanleitung, Anzeigemenüs, Berichtserstellung und Batteriewechsel.
Vorview Elitech RC-17/RC-17N Einweg-Temperaturrekorder – Technische Daten und Anleitung
Detaillierte technische Spezifikationen, Funktionen und Gebrauchsanweisungen für die Einweg-Temperaturrekorder Elitech RC-17 und RC-17N, die zur Überwachung der Temperatur während des Transports und der Lagerung empfindlicher Güter entwickelt wurden.